Project | Model Performance parameters |
VICTOR 25
|
Source |
DC millivolt
| Range | -10 ~ 110mV |
Accuracy= ±(% of the set value+% of range) | 0.02%+0.01% |
Resolution | 1uV |
DC voltage
| Range | -0.1 ~ 11V |
Accuracy= ±(% of the set value+% of range) | 0.02%+0.01% |
Resolution | 10uV/0.1mV |
DC current
| Range | 0 ~ 22mA |
Accuracy= ±(% of the set value+% of range) | 0.02%+0.02% |
Resolution | 1uA |
Sink current
| Range | 0 ~ 22mA |
Accuracy= ±(% of the set value+% of range) | 0.02%+0.02% |
Resolution | 1uA |
External power supply | 5 ~ 28V |
Resistance
| Range | 0 ~40KΩ |
Accuracy= ±(% of the set value+% of range) | 0.02%+0.02% |
Resolution | 0.01Ω/0.1Ω/1Ω |
Thermocouple
| Range | R/S/K/E/J/T/B/N |
Accuracy | 0.4℃ |
Resolution | 0.1℃/1℃ |
RTD
| Range | Pt100/Pt200/Pt500/ Pt1000/Cu10/Cu50 |
Accuracy | 0.2℃ |
Resolution | 0.1℃ |
Frequency
| Range | 1Hz ~110KHz |
Accuracy | ±2words |
Resolution | 0.01Hz/1Hz/0.1KHz/2KHz |
Pulse
| Frequency | 1Hz ~10KHz |
Range | 1 ~100000cyc |
Accuracy | ±2words |
Resolution | 1cyc |
Switch
| Range | 1Hz ~110KHz |
Accuracy | ±2words |
Resolution | 0.01Hz/1Hz/0.1KHz/2KHz |
Pressure
| Range | Depends on the pressure module |
Accuracy= ±(% of the set value+% of range) | Depends on the pressure module |